Scintillation Photon Counting Detectors for 100 ps Time-of-Flight PET Imaging
Project Number5R01EB033402-03
Contact PI/Project LeaderCATES, JOSHUA WILLIAM
Awardee OrganizationUNIVERSITY OF CALIF-LAWRENC BERKELEY LAB
Description
Abstract Text
Project Summary
Clinical time-of-flight positron emission tomography (TOF-PET) systems capable of excellent coincidence time
resolution (CTR) promise to drastically enhance effective 511 keV photon sensitivity. The ability to more precisely
localize annihilation origins along system response lines constrains event data, providing improved signal-to-
noise ratio (SNR) and reconstructed image quality by associating 511 keV photons more closely to their true
origin. This SNR enhancement increases as CTR is improved, and a major goal of ongoing PET instrumentation
research and development is to push system CTR ≤100 ps full-width-at-half-maximum (FWHM). At this level of
performance, events are constrained ≤1.5 cm, providing more than a five-fold increase in SNR relative to a
system with no TOF capability. Advanced systems capable of ≤100 ps FWHM CTR would effectively more than
double or quadruple the effective 511 keV system sensitivity, in comparison to state-of-the-art, clinical TOF-PET
systems (250-400 ps FWHM CTR). Thus, advancing CTR is also a pathway for greatly improved system
sensitivity without increasing detection volume and system cost. Standard PET detectors comprising segmented
arrays of high-aspect-ratio scintillation crystal elements cannot achieve this level of performance and are
ultimately limited by poor light collection efficiency and depth-dependent scintillation photon transit time jitter
seen by the photodetector. To address this, we propose to develop a new detector readout concept which allows
scintillation photons to be counted and a unique timestamp to be assigned for the first arriving photon at each
photosensor pixel. We will leverage this new advancement in scalable PET detector readout and produce PET
detector modules capable of high resolution, three-dimensional positioning capabilities and 100 ps FWHM CTR
in a design that also makes no sacrifices on 511 keV photon detection efficiency. The new detector design will
be integrated into large area detector modules that span the full axial extent (>20 cm) of a clinical PET system,
including front-end signal and back-end data processing. We will construct a prototype tomographic imaging
setup and quantify relevant system performance metrics and the imaging performance of future clinical systems
made from this new detector. The proposed PET detector technologies can have a significant impact on
quantitative PET imaging. The image SNR enabled by the significant boost in effective sensitivity can be
employed to substantially reduce tracer dose and shorten scan time/increase patient throughput, or to better
visualize and quantify smaller lesions/features in the presence of significant background, which are important
features that can make PET more practical and accurate, as well as help to expand its roles in patient
management.
Public Health Relevance Statement
Project Narrative
We propose to develop and demonstrate the clinical imaging performance of high resolution, clinical time-of-
flight positron emission tomography detectors that achieve 100 ps FWHM coincidence time resolution. This
detector technology can facilitate PET systems with substantially enhanced reconstructed image signal-to-noise
ratio and contrast-to-noise ratio, which can be exploited to improve lesion detection and quantification, or greatly
reduce injected radiation dose or scan duration, making PET more accurate, safe, and practical across a wide
variety of imaging studies.
National Institute of Biomedical Imaging and Bioengineering
CFDA Code
286
DUNS Number
078576738
UEI
ENBLDJUN4N73
Project Start Date
30-September-2022
Project End Date
30-June-2026
Budget Start Date
01-July-2024
Budget End Date
30-June-2025
Project Funding Information for 2024
Total Funding
$755,960
Direct Costs
$497,702
Indirect Costs
$258,258
Year
Funding IC
FY Total Cost by IC
2024
National Institute of Biomedical Imaging and Bioengineering
$755,960
Year
Funding IC
FY Total Cost by IC
Sub Projects
No Sub Projects information available for 5R01EB033402-03
Publications
Publications are associated with projects, but cannot be identified with any particular year of the project or fiscal year of funding. This is due to the continuous and cumulative nature of knowledge generation across the life of a project and the sometimes long and variable publishing timeline. Similarly, for multi-component projects, publications are associated with the parent core project and not with individual sub-projects.
No Publications available for 5R01EB033402-03
Patents
No Patents information available for 5R01EB033402-03
Outcomes
The Project Outcomes shown here are displayed verbatim as submitted by the Principal Investigator (PI) for this award. Any opinions, findings, and conclusions or recommendations expressed are those of the PI and do not necessarily reflect the views of the National Institutes of Health. NIH has not endorsed the content below.
No Outcomes available for 5R01EB033402-03
Clinical Studies
No Clinical Studies information available for 5R01EB033402-03
News and More
Related News Releases
No news release information available for 5R01EB033402-03
History
No Historical information available for 5R01EB033402-03
Similar Projects